SMITH & NEPHEW MEDICAL LTD. RENASYS TOUCH C2 POWER SUPPLY; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801286
Device Problem Electrical Shorting (2926)
Patient Problem Electric Shock (2554)
Event Date 09/20/2018
Event Type  malfunction  
Event Description
It was reported that a transformer block has the wires exposed and a patient received an electrical shock.A medical intervention was not required.
 
Manufacturer Narrative
As of today, no sample or additional information requested for this complaint has become available.Without further information we cannot progress our investigation or confirm the details supplied in this complaint.In the absence of additional information, our investigation remains inconclusive.At this time an exact root cause cannot be determined.If additional information becomes available in the future, this case will be reopened.
 
Manufacturer Narrative
Our investigation into the complaint has now concluded.The device was sent to our technical centre so a thorough technical analysis could be carried out to help to identify if there were any problems and/or what could have caused the reason for the complaint.A visual inspection of the device found that the cable strain reliefs were damaged, however the reported malfunction was not observed during a functional evaluation as although internal wires were exposed the insulation on these was fully intact hence it is not possible for the patient to receive an electric shock from the device.Following the complaint device evaluation it was recommended that due to the exposed insulation in the power cord the device should be scrapped.Smith and nephew are continually investigating ways to develop and improve our products and we will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TOUCH C2 POWER SUPPLY
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull, HU3 2 8N
UK  HU3 28N
MDR Report Key7980923
MDR Text Key124350604
Report Number8043484-2018-00333
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 01/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number66801286
Date Manufacturer Received09/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-