GUIDANT CRM CLONMEL IRELAND TELIGEN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number E102
Device Problems Over-Sensing (1438); Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574); Ambient Noise Problem (2877)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 08/02/2018
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that during an esophagogastroduodenoscopy, the patient received multiple shocks and anti-tachycardia pacing (atp) therapy due to oversensing of noise from electrocautery.There was also pacing inhibition with longer durations due to the oversensing.It was noted that a magnet had been placed as a precaution for the electrocautery usage.The implantable cardioverter defibrillator's (icd) memory and the other manufacturer's magnet were sent into boston scientific engineering for review.Engineering tested the magnet with a gauss meter and found that it exceeded the minimum requirements for magnetic field strength.Review of the icd's memory found 61 recorded magnet applications.Engineering noted that the magnet may have been slipping off during the procedure.The icd and rv lead remain in service and no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELIGEN
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7981279
MDR Text Key124337507
Report Number2124215-2018-17346
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/18/2012
Device Model NumberE102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received08/02/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/23/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
-
-