GUIDANT CRM CLONMEL IRELAND ACCOLADE IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problems Ambient Noise Problem (2877); Device Sensing Problem (2917); Impedance Problem (2950)
Patient Problem Ventricular Tachycardia (2132)
Event Date 03/30/2018
Event Type  malfunction  
Manufacturer Narrative
A boston scientific technical services consultant reviewed data and identified a respiratory rate trend (rrt) signal that was present but did not result in any oversensing. The consultant instructed the caller to evaluate atrial sensing and perform isometrics and pocket manipulation while sample impedance measurements and program off rrt. It was noted that the patient was in ventricular tachycardia (vt) at the end of episode. The consultant suggested lowering the vt trigger rate. As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this system exhibited small amplitudes and jumpy impedance measurements on the atrial channel. No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACCOLADE
Type of DeviceIMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7981898
MDR Text Key124337184
Report Number2124215-2018-15591
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/10/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/21/2018
Device Model NumberU128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received08/10/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-