Boston scientific received information that two weeks post implant this pacemaker measured increased right ventricular (rv) thresholds.The outputs were increased to insure an adequate safety margin.Five weeks later, the patient was seen again and the rv threshold values had again increased.The physician elected to reposition the rv lead.Following the rv lead revision procedure, this device had an estimated longevity of 2 years.A request was made to have data from this device analyzed.Data analysis showed the battery appeared to be depleting more quickly than expected.Device replacement was recommended.The device was explanted without incident.
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.The titanium case was opened and visual inspection revealed no irregularities.Testing found that the battery had depleted earlier than expected.When connected to an external power source, the device passed all tests and operated normally.No high current drain or other device anomaly was identified during analysis.Although the battery depleted early, laboratory analysis was unable to reproduce the condition that caused the battery to deplete prematurely.
|