• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CPI - DEL CARIBE INGEVITY; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CPI - DEL CARIBE INGEVITY; IMPLANTABLE LEAD Back to Search Results
Model Number 7741
Device Problems Over-Sensing (1438); Human-Device Interface Problem (2949)
Patient Problems Atrial Perforation (2511); No Known Impact Or Consequence To Patient (2692)
Event Date 09/04/2018
Event Type  Injury  
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
 
Event Description
Boston scientific received information that this right atrial (ra) lead was an attempted implant.During the implant procedure, the ra lead exhibited oversensing and was found to have perforated the heart wall.Blood was drained from the pericardium and a new ra lead was placed.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY
Type of Device
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7984385
MDR Text Key124316760
Report Number2124215-2018-18866
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/03/2020
Device Model Number7741
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received09/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
7736; 7741; 7742; L311
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age76 YR
-
-