• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND ACCOLADE IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND ACCOLADE IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problem Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/30/2018
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this caller was inquiring about data generated from this system that was suspected to be due to interaction with the minute ventilation (mv) feature. It was noted that this device does not have mv. A boston scientific technical services consultant discussed that the respiratory rate trend (rrt) feature uses the same signal as mv and instructed the caller to program this off. No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACCOLADE
Type of DeviceIMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7984542
MDR Text Key124340068
Report Number2124215-2018-18157
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date07/06/2019
Device Model NumberU128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received09/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/19/2018 Patient Sequence Number: 1
Treatment
4076; 4672; ADDRS1; U128
-
-