The light source was returned and evaluated; our findings were as follows: all output readings are below spec.
Unit was received in with a different lamp installed than what our records show we installed when we last serviced the light source in july 2016; also, the lamp differs from what karl storz recommends.
Lamp output is not meeting spec.
Lumen reading: 1113 actual; spec rating 1530-2800.
Optical power 1.
9 actual; spec rating 5.
1 max.
In addition, the model numbers for the light cable and arthroscope used for the procedure, which were provided to us by the hospital, are not of our manufacturer; the light cable, model #wa03310a, we believe is an olympus light cable; the arthroscope, model #502104030, is a stryker model number.
There was no reported malfunction of the light source.
The burn injury most likely occurred as a result of the light cable's connector coming into contact with the patient.
|