Model Number LXT
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Vitreous Floaters (1866); Discomfort (2330)
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that the presence of residual vitreous was observed in several patients eyes that were operated by different surgeons.The residual vitreous had the appearance of a filament, a long visible floating line.The patients experienced visual discomfort and were reoperated using fluid air exchange to eliminate the floating bodies.The exact number of patients and medical devices involved is unknown.This is one of four reports received from this facility.The manufacturer's reference numbers are: (b)(4).
 
Manufacturer Narrative
No sample has been returned for evaluation; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, lot history and complaint history reviews could not be conducted.A sample was not returned by the customer and no lot information was provided, therefore, the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown as no sample was returned; therefore, specific action with regards to this complaint cannot be taken.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7987020
MDR Text Key124469409
Report Number2028159-2018-02209
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberLXT
Device Catalogue Number8065751145
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/08/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-