• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE MAHWAH CARDIOSAVE SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CARDIOSAVE SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problems Loss of Power (1475); Computer Operating System Problem (2898)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/27/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date. A getinge field service engineer (fse) evaluated the iabp and it would not boot up in regular or service mode and f7 fault code was also observed. The fse replaced the front end board and performed 30psi calibration. Unit passed all calibration, functional and safety tests performed. Unit was returned to customer and cleared for clinical use. The full name of the initial reporter should read (b)(6), but the first name was abbreviated due to the contents exceeding character limit.
 
Event Description
It was reported that while in use on a patient, the cardiosave intra-aortic balloon pump (iabp) shut off and was unable to boot up. No patient injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7992069
MDR Text Key124796087
Report Number2249723-2018-01822
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation
Type of Report Initial
Report Date 10/22/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number0998-00-0800-XX
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-