The investigation determined that higher than expected vitros k+ and na+ results were obtained from a non-vitros biorad level 1 quality control fluid processed using vitros k+ and na+ slides in combination with a vitros 5600 integrated system and a vitros 4600 chemistry system.The assignable cause was most likely atypical calibrations.The calibrations were suboptimal when compared to typical calibration responses and parameters with responses from level 3 of the calibrator for both instruments being higher than the database.The issue was resolved by recalibrating with freshly reconstituted calibrator vials and using the same vitros na+ and vitros k+ reagent lots.The historical quality control results indicate both the vitros na+ lot and vitros k+ lot in combination with both the vitros 5600 and vitros 4600 systems were performing as expected.There is no indication of a reagent issue or an instrument issue.
|