BECTON DICKINSON, S.A. BD PHASEAL¿ INJECTOR LUER LOCK N35; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 515003
Device Problem Difficult or Delayed Separation (4044)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/03/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd phaseal¿ injector luer lock n35 injections would not disengage.There was no report of exposure, injury, or medical intervention.
 
Manufacturer Narrative
As no physical sample, picture sample, or lot number was provided for evaluation by our quality engineer team, a complete investigation could not be performed.Inspections and tests the tests performed during the manufacturing process to avoid faulty parts are listed below: during molding process (according ph-300 current version): visual inspections for injector parts (cylinder, needle housing, safety sleeve, piston and membrane) are performed by the operator to avoid faulty parts (flashes, unfilled and burned parts, etc).Critical to quality dimensions of all injector components are measured to check if the dimensions are within tolerance.Assembly process: (according to ph-301 current version) the following visual inspections are performed by the operator: safety sleeve must be connected and should be turning with the cylinder and piston.The functionality of the grips is verified.Verify the correct welding of the membrane, color and aspect.Cylinder assembly.Piston must be fixed by the safety sleeve.Needle housing should rotate clockwise and tip of the cannula must be observed.Cannula length (with a caliper gauge).Functionality and piston welding test: quality and functionality of the membrane is verified after be welding and punctured by the cannula.Retained samples cannot be taken for investigation as no lot was provided.No pictures or samples are available.The root cause cannot be established.
 
Event Description
It was reported that the bd phaseal injector luer lock n35 injections would not disengage.There was no report of exposure, injury, or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PHASEAL¿ INJECTOR LUER LOCK N35
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
BECTON DICKINSON, S.A.
camino de valdeolivia
s/n
san agustin de guadalix
MDR Report Key7996304
MDR Text Key124823216
Report Number3003152976-2018-00455
Device Sequence Number1
Product Code ONB
UDI-Device Identifier30382905150037
UDI-Public30382905150037
Combination Product (y/n)N
PMA/PMN Number
K140591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number515003
Device Lot NumberUNKNOWN
Date Manufacturer Received10/03/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-