The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 220 (pump max).During the procedure, the pump max was accidentally dropped on the floor and consequently, the power inlet became faulty; therefore, the pump max was disconnected.The procedure was completed successfully with a non-penumbra catheter and there was no adverse effect to the patient.
|
Results: the power inlet of the pump max was cracked and pushed in.The back of the pump max had a crack.The pump max had clear tape holding the pump together.After opening the pump max, it was noticed that the one of the cables was detached from the power inlet.The fuse box was also loose and would not stay in on its own.After reconnecting the detached cable and attempting to power on the device, there was no aspiration.Aspiration only occurred when the penumbra investigator applied pressure to the loose fuse box.When the pump max was powered on, it was able to achieve a vacuum pressure of approximately -28.0 in hg.Conclusions: evaluation of the returned pump max revealed a device with a damaged power inlet that was unable to produce sufficient vacuum pressure.The power inlet and device housing had cracks and the fuse box was loose from the power inlet.During functional testing, the pump max was unable to power on.Once the penumbra investigator reconnected a detached cable and applied pressure to the loose fusebox, the pump max was able to power on and produce a vacuum pressure within specification.The damage to the pump max was likely due to the reported dropping of the pump max onto the floor.Penumbra aspiration pumps are 100% visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|