Image review: cine image provided by the customer shows the cut end of the sheath in the anatomy.Product analysis summary: as received the sheath is cut in two approximately 1.5cm from the hub and the sheath is wrinkled and kinked, and the very distal edge/tip of the sheath is creased/damaged.None of the sheath is missing, full length is accounted for and none of the tip appears to be missing, just damaged.The cut in the sheath was more than ¾ way through, the remaining material that was still attached was stretched and pulled and broke, separating the proximal sheath from the hub.If information is provided in the future, a supplemental report will be issued.
|