• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND ACCOLADE IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND ACCOLADE IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U128
Device Problems High impedance (1291); Ambient Noise Problem (2877)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/21/2018
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy pacemaker (crt-p) declared a lead safety switch (lss) due to high out of range pacing impedances on the right ventricular (rv) lead, measuring greater than 2000 ohms. The patient's rv lead was a non-boston scientific product. Noise was observed on the rv lead after the pace impedance spikes occurred. The rv lead configuration was reprogrammed, and the physician elected to continue to monitor the patient/system. This crt-p system remains in service. No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACCOLADE
Type of DeviceIMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key8008590
MDR Text Key125099991
Report Number2124215-2018-19201
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial
Report Date 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/15/2018
Device Model NumberU128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received09/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/15/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/26/2018 Patient Sequence Number: 1
Treatment
5076; MISMATCH; U128
-
-