TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500 BLOOD GAS MONITOR Back to Search Results
Model Number CDI510H
Device Problem Leak/Splash (1354)
Patient Problem Blood Loss (2597)
Event Date 10/02/2018
Event Type  Injury  
Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed. Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available. (b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, when the shunt sensor was attached to the bpm probe in the artery side of the bpm, leakage was observed. *there was blood loss of less than 5 cc, *product was changed out, *procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on october 26, 2018. The returned sample was visually inspected. There were no visual anomalies indicative of issues with the shunt body or large bore cap connection. The returned sample was then leak tested by connecting with the calibrated manometer, submerged into a water bath, and pressurized. A leak was noted at 1000 mmhg. The large bore adapter blue cap was then loosened and re-tightened by hand. The sample was then leak tested by connecting with the calibrated manometer, submerged into a water bath, and pressurized up to 1030 mmhg, a leak was noted from the large bore adapter cap at 1030 mmhg. A retention sample of the same product code and lot number combination was visually inspected, and no visual anomalies were noted. The retention sample was also leak tested up to 1030 mmhg, and no leaks were noted. The complaint was confirmed and the root cause for this event is a lack of sealing between the large bore adapter (blue cap) and the shunt sensor body. When the large blue vent cap is tightened, interference between the threads of the cap and the shunt sensor body exist, causing an inadequate seal leading to a leak. All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSHUNT SENSOR SYS500
Type of DeviceBLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
cathleen hargreaves
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key8011137
MDR Text Key125273240
Report Number1124841-2018-00269
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date03/31/2019
Device Model NumberCDI510H
Device Catalogue NumberN/A
Device Lot NumberWE30E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 10/28/2018 Patient Sequence Number: 1
-
-