DATASCOPE MAHWAH CS300; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem Failure to Calibrate (2440)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/03/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) is not required to be reviewed per getinge standard operating procedure since the iabp was manufactured more than a year before the date of event.A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.The fse checked the logs and also noticed an autofill failure.The autofill fault code pointed to possible leak or blood in the iabp circuit.The blood back tubing had a little moisture and was very yellow.The fse replaced the blood back detect tubing and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) displayed "unable to calibrate iab optical sensor".No patient harm, serious injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key8011302
MDR Text Key125604196
Report Number2249723-2018-01864
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-