• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC ANALYZER, PACEMAKER GENERATOR FUNCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Device Problem Device Sensing Problem (2917)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/11/2018
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that the mobile programmer exhibited a loss of sensing, during an implant procedure. A different model programmer was used to complete lead testing, without issue. The programmer remains in use. No patient complications have been reported as a result of this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARELINK SMARTSYNC
Type of DeviceANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key8013606
MDR Text Key125283277
Report Number3004593495-2018-01027
Device Sequence Number1
Product Code DTC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK163008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,FOREIG
Reporter Occupation
Type of Report Initial
Report Date 10/29/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/29/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number24970A
Device Catalogue Number24970A
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/11/2018
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-