• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - SEWING RING; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - SEWING RING; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1153
Device Problems Device Damaged Prior to Use (2284); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2018
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the "felt" on the ventricular assist device (vad) sewing ring was pulling away from the metal, resulting in blood leaking between the material and metal.The sewing ring was exchanged.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the sewing ring was not returned for evaluation.The reported event of sewing ring damage could not be confirmed due to insufficient evidence.Review of the inspection records confirmed that the sewing ring met all requirements prior to release.Applicable risk documentation and experience with events of similar circumstances were considered; events involving fabric de-laminating the ring are most often attributed to deviations from procedure during the manufacturing process.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - SEWING RING
Type of Device
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key8014614
MDR Text Key125463478
Report Number3007042319-2018-05199
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707002639
UDI-Public00888707002639
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2020
Device Model Number1153
Device Catalogue Number1153
Was Device Available for Evaluation? No
Date Manufacturer Received02/13/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age63 YR
-
-