• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BERND RAKOW HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BERND RAKOW HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 7050 USA#HL
Device Problem Pressure Problem (3012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/24/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).Maquet cardiopulmonary (b)(4) requested the product for investigation but the product was already discarded by the hospital.Therefore no laboratory investigation could be performed by the manufacturer.A review for similar complaints to be investigated already was performed and no similar complaints were found.Thus the reported failure could not be confirmed.Based on the information available at this time the cause of this failure was determined to not be attributed to a device related malfunction.Since the reported failure did not contribute to a death or serious injury and no systemic issue could be determined no corrective action is needed at this time.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
According to the hospital: va pt cannulated with 17french arterial and 25 french multi fenestrated venous cannula.The cardiohelp reads higher post mo (part) pressure than pre mo (pint).Which gave usa -44 reading.Also the venous return pressure (pven) does not match the cannula size to flow rpm.Rpm at 3900 with flow of only 4.5 lpm, only to have -20 venous pressure.They are swapping out hls sets anytime.No harm reported to the pt.Additional information received: customer switched out the hls circuit last night.Unfortunately the hls set was discarded.All three pressures were out of normal.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key8015158
MDR Text Key125605531
Report Number8010762-2018-00291
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/21/2020
Device Model NumberBEQ-HLS 7050 USA#HL
Device Catalogue Number701052794
Device Lot Number70125827
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/24/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-