Catalog Number PMX110
Device Problems Pumping Problem (3016); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/01/2018
Event Type  malfunction  
Manufacturer Narrative
Results: the vacuum pressure on the pump gauge and the calibrated gauge were within specification.The pump was opened by the penumbra investigator and there was no loose components inside.The pump max was plugged in and power on and functioned without an issue.Conclusions: evaluation of the returned pump max revealed that the device functioned without an issue.During the functional test, the pump max was plugged in and power on and functioned without an issue.The vacuum pressure on the pump gauge and the calibrated gauge were within specification.The pump was opened by the penumbra investigator and there was no loose components inside.The root cause of the reported vacuum and noise issue could not be determined.Penumbra pumps are visually inspected and 100% functionally tested during incoming by quality.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a medical procedure using a penumbra system aspiration pump max 110 (pump max).During the procedure, the physician observed that the pump max was aspirating irregularly.It was also reported that the pump max was making unusual noises.Therefore, the procedure was completed using another pump max.There was no adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key8015806
MDR Text Key125431988
Report Number3005168196-2018-02141
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,10/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPMX110
Device Lot NumberF16768-03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received10/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/04/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-