MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST SINGLE DETECTOR; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST SINGLE DETECTOR; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER) Back to Search Results
Model Number 712020
Device Problem Defective Device (2588)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The investigation is still ongoing on this event.When the investigation is completed a follow-up report will be sent to the fda.
 
Event Description
The customer complained that one hinge of the foot board of the stitching patient support was defect.No injury occurred.
 
Manufacturer Narrative
Int.Ref.(b)(4).The digitaldiagnost is a direct digital radiography system with flat detector technology based on modular components to allow for customization for all radiographic applications and workload requirements.The patient support for stitching is an accessory that enables the examination of full leg or full spine, with the patient standing upright.For easy use, the patient support for stitching has wheels and a folding footboard.The footboard is connected to the frame via two hinges and has to be folded up and fixed by a hook for transportation, e.G.From one room to another.Philips field service engineer investigated at site and confirmed the reported issue.A rivet on one of the two hinges was broken and causing play in the movement of the affected hinge.If the both hinges of the footboard are for some reason worn differently, the footboard may folded up in a way that prevents the hook for snapping into place securely.Philips field service engineer replaced the two hinges and made sure that the equipment is working as specified.An already initiated field safety corrective action (fco (b)(4)) contains improvement of hinges and an additional brake cylinder.Correction: health professional instead of consumer result and conclusion.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIGITALDIAGNOST SINGLE DETECTOR
Type of Device
SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstr. 24
hamburg 22335
GM  22335
MDR Report Key8018570
MDR Text Key125613114
Report Number3003768251-2018-00009
Device Sequence Number1
Product Code MQB
Combination Product (y/n)N
PMA/PMN Number
K090625
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Remedial Action Recall
Type of Report Initial,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number712020
Device Catalogue Number712020
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2863-2018
Patient Sequence Number1
-
-