The sample evaluation confirms the reported condition of breach of sterility.
A slice was found to be present in the outer carton and when the outer carton was opened the slice was found to have gone through the lid of the tyvek tray which contained the perfix plug device, confirming a breach of sterility.
No manufacuturing anomalies were found.
The slice was likely caused by a sharp object, i.
E.
Box cutter or other type of blade, used during the removal from a shipper box that penetrated through the packaging.
When the damage occurred cannot be determined.
To date this is the only reported complaint for this production lot of (b)(4) units released for distribution in september, 2018.
A review of the manufacturing records was performed and found that the lot was manufactured to specification.
There was no patient involvement.
|