MAQUET SAS HANAULUX; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Catalog Number 567910910
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2018
Event Type  malfunction  
Manufacturer Narrative
We are aware that this is past the 30 day deadline for reporting.We have reminded the complaint handler to review fda reporting guidelines in order to prevent this from happening again the issue is being investigated by manufacturing site.(b)(4).
 
Event Description
On (b)(6) 2018 maquet sas became aware of an incident with one of surgical lights- hanaulux 3000.As it was stated, the spring arm weld is cracked.There is no injury reported however we decided to report the issue in abundance of caution as any part falling might cause an injury.Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
(b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer reference number: (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still investigated by manufacturing site.(b)(4).Exemption # e2018005.(b)(4).
 
Event Description
Manufacturer's reference number (b)(4).
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
The issue is still being investigated by manufacturing site.Getinge usa sales, llc (importer) is submitting this report on behalf of the legal manufacturer of the device maquet sas, parc de limère, avenue de la pomme de pi orléans cedex 2, france 45074 exemption # e2018005 getinge usa sales, llc 45 barbour pond drive wayne, nj 07470 contact person: (b)(4).
 
Manufacturer Narrative
On(b)(6)2018 maquet sas became aware of an incident with one of surgical lights- hanaulux 3000.As it was stated, the spring arm weld is cracked.There is no injury reported however we decided to report the issue in abundance of caution as any part falling might cause an injury.It was established that when the event occurred, the surgical light did not meet the manufacturer¿s specification.During the event occurrence, the device was not being used for patient treatment.During this investigation it was established that the reported issue has resulted in a serious injury or worse.The issue was reviewed in details by the supplier of the spring arm.The front joint was found broken on both sides.Both fractures are beside the area of welding seam.The result shows that the defect is not a welding seam break but a material breach.The subject matter experts of the manufacturing site concluded that the breaking surface in the upper position demonstrates a crack initiation caused by an overstressing in the past.The dull finish area of fracture shows a recurrence overstressing that led to the complete breakage over time.According to these facts, supplier assumed that the defect were a result of improper use.
 
Event Description
Manufacturer's reference number 2018-60066.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANAULUX
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
orléans cedex 2
FR 
MDR Report Key8020390
MDR Text Key125797814
Report Number9710055-2018-00152
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Replace
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 01/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number567910910
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-