• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP 5; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-01-04
Device Problems Use of Device Problem (1670); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/02/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was not provided.Livanova (b)(4) manufactures the centrifugal pump 5.The incident occurred in (b)(6).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue.The technician identified that fluid ingress shorted the centrifugal pump motor which led to the reported malfunction.The technician replaced the motor to resolve the reported issue.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a centrifugal pump 5 displayed an error message and could not be controlled during procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP 5
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key8023284
MDR Text Key125920029
Report Number9611109-2018-01310
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 10/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-01-04
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/02/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/08/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-