Catalog Number 70-6529
Device Problem Complete Loss of Power (4015)
Patient Problem No Patient Involvement (2645)
Event Date 04/30/2013
Event Type  malfunction  
Manufacturer Narrative
The leadcare ii blood lead analyzer and adapter was returned to (b)(4) for evaluation.During investigational testing, the returned unit was unable to power on and the returned adapter was able to power on multiple in-house units, therefore the failure is attributed to the analyzer.Because the customer indicated a burning smell and the result was total power loss, this is being reported because it has the potential for thermal injury.No report of injuries or erroneous results from the user facility.(b)(4).
 
Event Description
User facility contacted (b)(4) product support team to report a burning electrical smell coming from their leadcare ii blood lead analyzer when plugged into the a/c adapter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEADCARE II BLOOD LEAD TEST SYSTEM
Type of Device
LEADCARE II ANALYZER
Manufacturer (Section D)
MAGELLAN DIAGNOSTICS, INC.
101 billerica ave
building 4
n. billerica MA 01862
Manufacturer (Section G)
MAGELLAN DIAGNOSTICS, INC.
101 billerica ave
building 4
n. billerica MA 01862
Manufacturer Contact
heather reed
101 billerica ave.
building 4
n. billerica, MA 01862
9782484811
MDR Report Key8027087
MDR Text Key126382357
Report Number1218996-2018-00032
Device Sequence Number1
Product Code DOF
UDI-Device Identifier00850355006017
UDI-Public00850355006017
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052549
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number70-6529
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/28/2013
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/30/2013
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-