Catalog Number 71631066 |
Device Problems
Loose or Intermittent Connection (1371); Connection Problem (2900)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/05/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that a malfunction occured during operation, when the driver was connected to a proximal screw and rotated, the connection was not tight but loose.Surgeon had to divert from the original procedure using sureshot driver to continue the operation.There was no delay and no patient harm due to this issue.
|
|
Manufacturer Narrative
|
The medium hex driver was returned and evaluated.A visual inspection found that the hex end of the driver had deformation from use.The threads on the driver shaft were worn and would not allow the corresponding screw to properly seat; this was confirmed with a function check.We recommend that all reusable instruments be routinely inspected for wear and damage so they can be replaced as necessary.No additional complaints were found during the complaint history review.Reviewing the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Based on this investigation, the need for corrective action is not indicated.No additional actions are being taken at this time; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
|
|