• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MEDIUM HEXDRIVER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MEDIUM HEXDRIVER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71631066
Device Problems Loose or Intermittent Connection (1371); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2018
Event Type  malfunction  
Event Description
It was reported that a malfunction occured during operation, when the driver was connected to a proximal screw and rotated, the connection was not tight but loose.Surgeon had to divert from the original procedure using sureshot driver to continue the operation.There was no delay and no patient harm due to this issue.
 
Manufacturer Narrative
The medium hex driver was returned and evaluated.A visual inspection found that the hex end of the driver had deformation from use.The threads on the driver shaft were worn and would not allow the corresponding screw to properly seat; this was confirmed with a function check.We recommend that all reusable instruments be routinely inspected for wear and damage so they can be replaced as necessary.No additional complaints were found during the complaint history review.Reviewing the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Based on this investigation, the need for corrective action is not indicated.No additional actions are being taken at this time; however we will continue to monitor for future complaints and investigate further as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDIUM HEXDRIVER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8028827
MDR Text Key125819616
Report Number1020279-2018-02329
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71631066
Device Lot Number08AM09167
Date Manufacturer Received10/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-