Model Number MMT-751LNAS
Device Problems Fluid Leak (1250); Priming Problem (4040)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/17/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned. No conclusion can be drawn at this time. We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call that the insulin pump had prime fill anomaly. The customer¿s blood glucose level was unknown at the time of incident. The customer state that insulin squirts during priming. The insulin pump will not be returned for the analysis.
 
Manufacturer Narrative
Insulin pump passed the displacement, rewind, basic occlusion, occlusion, prime and excessive no delivery test. No excessive no delivery alarm, rewind anomaly and prime/fill anomaly noted during test. (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name530G INSULIN PUMP MMT-751LNAS
Type of DeviceARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key8036882
MDR Text Key126137579
Report Number2032227-2018-55723
Device Sequence Number0
Product Code OZO
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/22/2019
2 Devices were Involved in the Event: 1   2  
0 Patients were Involved in the Event:
Date FDA Received11/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMMT-751LNAS
Device Catalogue NumberMMT-751LNAS
Device Lot NumberA5751LNASJ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/21/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-