RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Output Problem (3005); Temperature Problem (3022)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2018.Date of report:(b)(6) 2018.Philip's field service engineer (fse) evaluated the device and could not duplicate the reported malfunction however, the error code was verified in the event log.To address the issue, the fse replaced the blower assembly per the service manual.The ventilator passed all testing and operated within the manufacturing specifications.
|
|
Event Description
|
The customer reported that during clinical use, the ventilator generated an error relevant to high blower temperature.The ventilator was in clinical use at the time of the event.No patient harm was reported.The event date was not specified; estimate used.
|
|
Manufacturer Narrative
|
Date of report: 16jan2019.Date rec'd by mfr: 14jan2019.The blower assembly was received at philips for evaluation.A visual inspection of the blower assembly outer surface revealed no evidence of damage or contamination.The blower assembly passed all testing.A high blower temperature alarm (1122 error code) could not be duplicated.There were no faults found with the blower assembly.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|