Model Number V60
Device Problems Output Problem (3005); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2018.Date of report:(b)(6) 2018.Philip's field service engineer (fse) evaluated the device and could not duplicate the reported malfunction however, the error code was verified in the event log.To address the issue, the fse replaced the blower assembly per the service manual.The ventilator passed all testing and operated within the manufacturing specifications.
 
Event Description
The customer reported that during clinical use, the ventilator generated an error relevant to high blower temperature.The ventilator was in clinical use at the time of the event.No patient harm was reported.The event date was not specified; estimate used.
 
Manufacturer Narrative
Date of report: 16jan2019.Date rec'd by mfr: 14jan2019.The blower assembly was received at philips for evaluation.A visual inspection of the blower assembly outer surface revealed no evidence of damage or contamination.The blower assembly passed all testing.A high blower temperature alarm (1122 error code) could not be duplicated.There were no faults found with the blower assembly.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key8037924
MDR Text Key127896854
Report Number2031642-2018-02376
Device Sequence Number0
Product Code MNT
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2018
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/28/2018
Was the Report Sent to FDA? No
Date Manufacturer Received10/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
PATIENT CIRCUIT, MASK, HUMIDIFIER: UNKNOWN
-
-