Model Number CYF-5
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2018
Event Type  Malfunction  
Manufacturer Narrative

The subject device in this report has not been returned to omsc for evaluation. The exact cause of the reported event could not be conclusively determined at this time. If additional information becomes available, this report will be supplemented.

 
Event Description

Olympus medical systems corp. (omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device. [first time; (b)(6) 2018]: kocuria rhizophila, micrococcus luteus, and staphylococcus epidermidis (2 cfu / 100ml). [second time; (b)(6) 2018]: staphylococcus warneri (1 cfu / 100ml). The device had been reprocessed using peracetic acid. There was no report of infection associated with this report.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFIBERSCOPE "CYF-5", EUROPEAN VERSION
Type of DeviceCYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key8049862
MDR Text Key129050071
Report Number8010047-2018-02142
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type FOREIGN,HEALTH PROFESSIONAL,U
Reporter Occupation BIOMEDICAL ENGINEER
Type of Report Initial,Followup,Followup
Report Date 01/16/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received11/08/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL NumberCYF-5
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/18/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured03/13/2014
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-