• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. FIBERSCOPE "CYF-5", EUROPEAN VERSION; CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. FIBERSCOPE "CYF-5", EUROPEAN VERSION; CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/08/2018
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device.[first time; (b)(6) 2018]: kocuria rhizophila, micrococcus luteus, and staphylococcus epidermidis (2 cfu / 100ml).[second time; (b)(6) 2018]: staphylococcus warneri (1 cfu / 100ml).The device had been reprocessed using peracetic acid.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) reviewed the manufacture history of the device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) but was returned to olympus (b)(4).(b)(4) sent the subject device to a third party laboratory for microbiological testing.As a result of the testing, no microbe was detected from the sample collected from all channels of the subject device.The testing result cleared the (b)(4) guideline.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERSCOPE "CYF-5", EUROPEAN VERSION
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to
MDR Report Key8049862
MDR Text Key129050071
Report Number8010047-2018-02142
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup,Followup
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/18/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-