BD MEDICAL - DIABETES CARE BD ULTRA-FINE¿ INSULIN SYRINGE Back to Search Results
Catalog Number 320440
Device Problem Difficult to Advance (2920)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/19/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: a complaint history check was performed and this is the 1st related complaint for difficult to operate on lot # 7177921.Customer returned (93) 3/10cc, 8mm, 31g syringes (13 in open poly bags, 80 in sealed poly bags) with the shelf carton from lot # 7177921.Customer states that it was difficult to aspirate into the syringe.Thirty out of 93 returned syringes were tested (13 from open poly bags, 17 from the sealed poly bags) and all were able to draw and expel properly without any observed defects.A review of the device history record was completed for batch # 7177921 all inspections were performed per the applicable operations qc specifications.There were zero (0) notifications noted that pertained to the complaint.Investigation conclusion: bd was not able to duplicate or confirm the customer¿s indicated failure complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.Root cause description: root cause cannot be determined at this time as the issue is unconfirmed.Rationale: based on the investigation, no additional investigation and no capa is required at this time.
 
Event Description
It was reported that bd ultra-fine¿ insulin syringe was difficult to operate or had restricted movement.No serious injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ULTRA-FINE¿ INSULIN SYRINGE
Type of Device
INSULIN SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key8052367
MDR Text Key128388057
Report Number1920898-2018-00849
Device Sequence Number1
Product Code FMF
UDI-Device Identifier00382903204403
UDI-Public382903204403
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2022
Device Catalogue Number320440
Device Lot Number7177921
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-