• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL REVEAL LINQ DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL REVEAL LINQ DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problems Premature Discharge of Battery (1057); False Positive Result (1227); Incorrect Measurement (1383); Over-Sensing (1438); Unable to Obtain Readings (1516); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/13/2017
Event Type  Malfunction  
Manufacturer Narrative

This device was reported as included in the field action. Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action. It is included in the field action in the abundance of caution. If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that the implantable cardiac monitor (icm) experienced right ventricular (rv) oversensing during an atrial fibrillation (af) episode and the device was showing invalid histograms. It was noted that the device was at end of service (eos). It was further reported that the icm had not received the software update for elective replacement indicator (eri) trigger and was displaying early eri. This was resolved once the device software was updated. The icm remains in use. No patient complications have been reported as a result of this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREVEAL LINQ
Type of DeviceDETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH 1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key8059097
MDR Text Key127267968
Report Number9614453-2018-03917
Device Sequence Number1
Product Code DSI
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation
Remedial Action Recall
Type of Report Initial
Report Date 11/12/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received11/12/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator LAY USER/PATIENT
Device EXPIRATION Date02/28/2017
Device MODEL NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/19/2018
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured03/02/2016
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1249-2016

-
-