COOK INC ADVANCE 18 LP LOW PROFILE BALLOON CATHETER LIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number PTA4-18-150-5-20
Device Problem Fluid Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/12/2018
Event Type  malfunction  
Manufacturer Narrative
Pma/510(k) number: k130293. This report includes information known at this time. A follow-up report will be submitted should additional relevant information become available.
 
Event Description
It was reported, an (b)(6) male weighing (b)(6) was undergoing a left popliteal intervention procedure. On initial inflation, the advance 18 lp low profile balloon catheter developed a pinhole leak at the proximal end. The complaint device was inflated to 8 atmospheres (atm) with a mix of unspecified contrast to saline in a 4:10 ratio using a cook sphere inflation device. Length of inflation time was not known. There was no angulation but some calcification of the target vessel. The complaint device was removed and another device was used successfully to complete the procedure. A section of the device did not remain inside the patient¿s body. The patient did not require any additional procedures due to this occurrence. According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameADVANCE 18 LP LOW PROFILE BALLOON CATHETER
Type of DeviceLIT CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key8068905
MDR Text Key128253338
Report Number1820334-2018-03348
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/25/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/17/2021
Device Catalogue NumberPTA4-18-150-5-20
Device Lot Number9167081
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-