This is filed to report the thrombus and tissue damage.It was reported that this was a mitraclip procedure to treat grade 4 degenerative mitral regurgitation (mr).The steerable guide catheter (sgc) was inserted and the first clip (80222u118) was advanced to the mitral valve leaflets and positioned.During deployment when the actuator knob was pulled, the knob and the actuator mandrel came out of the system; however, the clip deployed successfully.At this point, a structure was visible on the tip of the sgc (could have been thrombus and/or tissue).The procedure continued and a second clip (80221u150) was positioned.During deployment when the actuator knob was pulled, the knob and actuator mandrel came out of the system.The clip still deployed successfully, reducing mr to 1.When the cds was removed, the thrombus or tissue disappeared.It was commented that the actuator knob may have been pulled too hard.The patient remained stable.There was no reported adverse patient sequela or a clinically significant delay during the procedure.No additional information was provided.
|
(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot.Based on information reviewed, a definitive cause for the reported thrombosis and tissue damage could not be determined in this incident.The reported patient effects of mitral valve tissue damage and thrombus are listed in the mitraclip system instructions for use as known possible complication associated with mitraclip procedures.Although a conclusive cause for the reported patient effects and the relationship to the device, if any, cannot be determined, there is no indication of a product quality issue with respect to manufacture, design or labeling.
|