• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS TCH DEVICE&POWER SPLY; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS TCH DEVICE&POWER SPLY; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Catalog Number 66801280
Device Problem Excessive Heating (4030)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/24/2018
Event Type  malfunction  
Event Description
It was reported that the pump renasys touch was in operation from (b)(6) 2018 to (b)(6) 2018 directly at patient.The pump overheats quickly.There was no impact to the patient.A back up was available, the treatment continued with the new device.
 
Manufacturer Narrative
The device was sent to our technical centre so an evaluation could be carried out to help to identify if there were any problems and/or what could have caused the reason for the complaint.A visual inspection was carried out and damages were observed on the outer casing.The reported complaints could not be replicated during the evaluation; a performance and safety check was conducted and no anomalies or problems were found, the device passed all functional tests and was confirmed working within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS TCH DEVICE&POWER SPLY
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
sarah freestone
101 hessle road
hull HU3 2-BN
UK   HU3 2BN
0447940038
MDR Report Key8073383
MDR Text Key127355882
Report Number8043484-2018-00386
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K153209
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66801280
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-