RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Overheating of Device (1437); Device Displays Incorrect Message (2591)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2018.Date of report: 15nov2018.The manufacturer's field service engineer (fse) performed remote troubleshooting with the customer.The customer verified the air inlet filter was clean and the cooling fan was running.The fse provided the customer with part information for the blower, software version 2.30, the manual reference for software installation and the service bulletin for software 2.30.The customer reported that the error did not recur.The air inlet filter was replaced and the respiratory therapist were reminded not to position the unit where the inlet is blocked.The unit was returned to service.The device not being used for treatment when the reported event occurred, and there is a relationship of the device to the reported problem.No parts were returned to failure investigation; therefore, the root cause of the reported issue could not be determined.
|
|
Event Description
|
It was reported that the unit logged an error 1122 (blower temperature high).The device was in use at the time of the event; however, there was no patient harm.The event date was not specified; estimate used.
|
|