Model Number V60
Device Problems Overheating of Device (1437); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2018.Date of report: 15nov2018.The manufacturer's field service engineer (fse) performed remote troubleshooting with the customer.The customer verified the air inlet filter was clean and the cooling fan was running.The fse provided the customer with part information for the blower, software version 2.30, the manual reference for software installation and the service bulletin for software 2.30.The customer reported that the error did not recur.The air inlet filter was replaced and the respiratory therapist were reminded not to position the unit where the inlet is blocked.The unit was returned to service.The device not being used for treatment when the reported event occurred, and there is a relationship of the device to the reported problem.No parts were returned to failure investigation; therefore, the root cause of the reported issue could not be determined.
 
Event Description
It was reported that the unit logged an error 1122 (blower temperature high).The device was in use at the time of the event; however, there was no patient harm.The event date was not specified; estimate used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RICA
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
lisa cardenas
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key8076807
MDR Text Key127408906
Report Number2031642-2018-02548
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/02/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/22/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
PATIENT CIRCUIT AND MASK: UNKNOWN
-
-