A steris service technician arrived onsite to inspect the sterilizer and found that during the last service activity the new o-ring was placed on top of the old one which, after several cycles, caused the new o-ring to become dislodged.As the o-ring became dislodged this allowed the valve to become stuck in the "open" position and steam to leak from the unit.The technician rebuilt the s2 valve and seated the o-ring, tested the unit, confirmed it to be operating according to specification, and returned it to service.The technician was retrained on proper maintenance practices.No additional issues have been reported.
|