BARD ACCESS SYSTEMS NAVARRE OPTI-DRAIN MULTI-USE DRAINAGE CATHETER 10F LOCKING PIGTAIL STAND, INFUSION Back to Search Results
Catalog Number NOD10LPT
Device Problem Crack (1135)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/03/2018
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed. The sample was not returned to the manufacturer for inspection/evaluation. Therefore, the investigation of the reported event is inconclusive. Based upon the available information, the definitive root cause for this event is unknown. The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device. Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
 
Event Description
It was reported that the drain was placed under ct. The day after placement the device was flushed as prescribed. When attempting to flush, allegedly it was noticed that the piece that connects the catheter to the drainage bag was split. The device was removed and replaced. There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNAVARRE OPTI-DRAIN MULTI-USE DRAINAGE CATHETER 10F LOCKING PIGTAIL
Type of DeviceSTAND, INFUSION
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1625 w 3rd st.
tempe, AZ 85281
4803032689
MDR Report Key8081967
MDR Text Key127541822
Report Number3006260740-2018-03267
Device Sequence Number1
Product Code FOX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 11/16/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberNOD10LPT
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-