Model Number MSERIES BI-PHASIC |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
Malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the device and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device was unable to obtain an ecg signal via electrode pads.
Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation evaluated the device and the reported malfunction was not replicated or confirmed.
The device was put through extensive testing which included bench handling and full ecg functionality testing using the returned multifunctional cable without duplicating the malfunction.
The device also passed the patient impedence calibration test.
The device was recertified and returned to the customer.
No trend is associated with reports of this type.
|
|