The user facility has stated that the device will remain at their site and will not return to olympus for evaluation.
The cause of the complaint cannot be confirmed.
A review of the device service history shows no olympus service since 2016.
Despite followup attempts by phone and in writing, no further information is available regarding the device malfunction, mechanism of air embolism injury by the device, or type of injury.
It has not been confirmed that the event involved any report of over-inflation by the device.
There is also no further information on the current status of the patient.
As a preventive measure against device malfunction, the instructions manual states that the device should be functionally inspected prior to the procedure, along with the other devices used.
The inspection includes a verification of airflow function and verification of airflow indicators on the device.
The instructions manual also warns that during the procedure, ¿anytime you observe an irregularity in a light source function, stop the examination immediately.
¿ the instructions manual notes, air and water is not fed even if the airflow indicator ¿on¿ is lighting when no endoscope is connected to the light source.
¿ and that the air/water function ¿is available only when a compatible fiber endoscope or videoscope is connected.
¿ the instructions manual also states, ¿in case of light source failure or malfunction, always keep another light source in the room ready for use.
¿.
|