(b)(4).Device evaluation by manufacturer: field service engineering (fse) was at customer's site to address the reported event.Customer confirmed low pressure error was still occurring.Upon troubleshooting, fse found that the solenoid valve was failing and replaced it.The customer successfully completed quality control run without any errors.No further action required by field service.The g8 instrument is functioning as expected.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4).There were no similar complaints identified during the search period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.101 pressure low: the pressure will not rise because the pump is unable to run due to air bubbles in the pump check valve.If the elution buffer is empty, place a new elution buffer and execute reagent change.Next, execute drain flush.See "chapter 5 section 5.5: pump air removal".Execute manual pumping using the pump key in the main screen (second screen), and open and close the drain valve 2 or 3 times.If the pressure rises when the drain valve is closed, the operation is complete.If the pressure still does not rise or stabilize, execute drain flush again.In addition, confirm that the drain valve is securely closed.The most probable cause of the reported event is pending investigation completion.
|
A customer reported getting sporadic 101 pressure low errors on the g8 instrument.The customer rebooted instrument and performed drain flush which got pressure in range, but dropped to zero after a few sample run.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
Tosoh bioscience, inc.Is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number e2017013.Submission of this report does not constitute an admission that the importer or manufacturer's product caused or contributed to the event.Device evaluation by manufacturer: the 2-way solenoid valve was returned to tosoh instrument service center for investigation.Functional testing could not confirm reported event.Part will be discarded.The most probable cause of the reported event is unknown; error was not duplicated.
|