MAQUET CV HS III PROXIMAL SEAL CLAMP, VASCULAR Back to Search Results
Catalog Number C-HS-3045
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4). A lot history record review was completed for the reported product lot number. There were no ncmr¿s for the reported lot number. Since the device is not available to be returned to us, a technical evaluation cannot be performed. Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Event Description
The hospital reported that during preparation for a coronary artery bypass procedure, hs iii proximal seal dropped into the loader when the delivery system was pulled out. The hospital tried to insert it after rolling the seal up by hand, but couldn't. A replacement device was used to complete the procedure. The hospital did not report any patient effects.
 
Manufacturer Narrative
Internal complaint (b)(4). Autonumber # (b)(4).
 
Event Description
The hospital reported that during preparation for a coronary artery bypass procedure, hs iii proximal seal dropped into the loader when the delivery system was pulled out. The hospital tried to insert it after rolling the seal up by hand, but couldn't. A replacement device was used to complete the procedure. The hospital did not report any patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHS III PROXIMAL SEAL
Type of DeviceCLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
MDR Report Key8095682
Report Number2242352-2018-01194
Device Sequence Number1
Product Code DXC
Combination Product (y/n)N
PMA/PMN Number
K130382
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/25/2019
Device Catalogue NumberC-HS-3045
Device Lot Number25138618
Was Device Available for Evaluation? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/28/2018
Is This a Reprocessed and Reused Single-Use Device? No
-
-