Catalog Number 284580
Device Problem Suction Failure (4039)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
If additional information should become available, a supplemental medwatch will be submitted accordingly. (b)(4). Investigation summary: the complaint device was received and evaluated at the service center. On inspection, the mother board was defective and the suction motor had an issue. The complaint is confirmed. The customer chose not to repair the device. The root cause for the device not working was a defective mother board and an issue with the suction motor. A review of the device history record indicated that this batch of product was processed without incident; therefore there is no evidence of manufacturing anomalies on the records reviewed. At this time, no corrective action is required, and no further action is warranted. However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the affiliates in (b)(6) that before surgery, it was observed that the 4580 fms duo+ pump/shaver combo device did not work. During in-house engineering evaluation, it was determined that the device had a suction issue. This event did not occur during surgery. There was no patient involvement. There were no reports of injuries, medical intervention or prolonged hospitalization. The exact date of the event was unknown but was noted to have occurred in 2018. All available information has been disclosed. If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name4580 FMS DUO+ PUMP/SHAVER COMBO -NS
Type of DeviceDISTENSION UNIT, FLUID, ARTHROSCOPIC
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8096307
MDR Text Key128197971
Report Number1221934-2018-55365
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 11/21/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number284580
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/21/2018
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured04/30/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-