• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND HAWKONE 6FR; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Catalog Number H1-M
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/19/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The physician intended to use a hawkone to treat a severely calcified 150mm lesion, with 90% stenosis in the right distal superficial femoral artery (sfa).The artery diameter was reported as 6 mm.A 6 fr sheath, a 0.014¿ guidewire and a 6mm spiderfx were used.Ifu was followed.After making one pass, the device thumb switch would not push forward to close the cutter window.The defective device was successfully removed without difficulty.Another hawkone was used and the patient was treated successfully.No patient injury was reported.
 
Manufacturer Narrative
The thumbswitch stopped functioning while in use in the patient.The cutter was outside the housing when it crashed.No deformation was noted to the cutter.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAWKONE 6FR
Type of Device
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key8099730
MDR Text Key128198626
Report Number9612164-2018-03334
Device Sequence Number1
Product Code MCW
UDI-Device Identifier00643169968332
UDI-Public00643169968332
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/16/2021
Device Catalogue NumberH1-M
Device Lot Number0009146284
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/28/2018
Date Device Manufactured05/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-