SMITH & NEPHEW, INC. INTERTAN 7.0MM COMP SCREW STARTER DRILL; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 71674570
Device Problem Break (1069)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Information (3190)
Event Date 10/12/2018
Event Type  malfunction  
Event Description
It was reported that a malfunction occurred during surgery.Drill was inserted down the drill guide.When the drill was activated the tip of the drill snapped off in patient.The surgeon was able to remove the broken piece and no further complications arose as a result of this.
 
Manufacturer Narrative
The affected intertan 7.00 compression starter drill was not returned for evaluation.A relationship, if any, between the subject affected device and the reported event could not be determined since the product was not returned for evaluation.Although the complainant indicated, the device was to be returned it has not been received by the manufacturer.As the device was not made available for investigation, the failure mode cannot be determined or confirmed.Additionally, the batch number was not provided; thus, a review of the manufacturing records cannot be conducted.Without the return of the actual product involved and no batch information available, our investigation of this report is inconclusive.Several potential factors that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to/after each use and cleaning.Additionally, failure to adhere to the warnings and precautions outlined in the instructions for use could result in damage and or impact the device functionality.There are no indications to suggest the device did not meet product specifications upon release into distribution.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 7.0MM COMP SCREW STARTER DRILL
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8105582
MDR Text Key128398246
Report Number1020279-2018-02382
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71674570
Device Lot NumberUNKNOWN
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-