MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problem Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/30/2018
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 8840, serial# unknown, product type: programmer, physician; product id: 8870, serial# unknown, product type: software.If information is provided in the future, a supplemental report will be issued.
 
Event Description
On (b)(6) 2018, information was received from an healthcare professional (hcp) regarding a patient receiving dilaudid (2 mg/ml, 0.6 mg/day) for non-malignant pain.On (b)(6) 2018, the rep was updating the patient's pump and the pump read memory error during the update.Upon re-interrogating the pump, the pump read that it had been programmed to stopped pump.The hcp was asking for assistance in troubleshooting.The hcp had been programming a bridge bolus (2 mg/ml to 10 mg/ml).The programming was reviewed and showed it had updated the new concentration and reservoir volume.The hcp updated the pump with the previous settings, then reprogrammed with the updated settings so that the bridge bolus would be accurate.The hcp confirmed on the call that the programming had resolved the issue and read accurately.No patient symptoms were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8108536
MDR Text Key129065679
Report Number3004209178-2018-26343
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169530119
UDI-Public00643169530119
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/14/2020
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/30/2018
Date Device Manufactured07/16/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
-
-