MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-20 |
Device Problem
Data Problem (3196)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/30/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8840, serial# unknown, product type: programmer, physician; product id: 8870, serial# unknown, product type: software.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
On (b)(6) 2018, information was received from an healthcare professional (hcp) regarding a patient receiving dilaudid (2 mg/ml, 0.6 mg/day) for non-malignant pain.On (b)(6) 2018, the rep was updating the patient's pump and the pump read memory error during the update.Upon re-interrogating the pump, the pump read that it had been programmed to stopped pump.The hcp was asking for assistance in troubleshooting.The hcp had been programming a bridge bolus (2 mg/ml to 10 mg/ml).The programming was reviewed and showed it had updated the new concentration and reservoir volume.The hcp updated the pump with the previous settings, then reprogrammed with the updated settings so that the bridge bolus would be accurate.The hcp confirmed on the call that the programming had resolved the issue and read accurately.No patient symptoms were reported.
|
|