Model Number EX1150
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
No complaint was received with return of the device. Failure event observed during analysis. During the analysis, the transmitter was presenting with a no power issue. Upon opening the transmitter, burnt marks were observed. The root cause of the no power issue was the burnt damage that the merlin. Net transmitter sustained through high current flow through the ac electrical power outlet.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERLIN@HOME, RF TELEMETRY BASIC
Type of DevicePACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8108765
MDR Text Key128564891
Report Number2017865-2018-17849
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial
Report Date 11/27/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberEX1150
Device Catalogue NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2017
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received11/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-