It was reported that this pacemaker system exhibited noise and oversensing on both the right atrial (ra) and right ventricular (rv) leads resulting in 4 seconds of asystole.The noise was able to be reproduced with isometrics in the bipolar configuration.The physician reprogrammed the device to fixed sensitivity, and the rv sensitivity was adjusted.Additionally, the programming for several of the blanking periods was adjusted.The patient had a remote follow up the following month and there were no new episodes containing noise.The physician plans to continue to monitor.This product remains in service.No adverse patient effects were reported.
|
The returned cardiac resynchronization therapy pacemaker (crt-p) was analyzed, and an engineering-level longevity prediction calculation was completed to assess the rate of battery depletion.Given the programmed parameters and other data stored within the memory of the device, the results of this calculation indicated that the actual rate of battery depletion fell within an acceptable range.Next, a series of diagnostic tests were conducted that verified the performance of pacing, sensing, defibrillation and recording functions.Having met the engineering longevity prediction, functionally passed all returned product testing, and with no further information to indicate a product performance issue, we have concluded that this device experienced normal battery depletion.
|
It was reported that this pacemaker system exhibited noise and oversensing on both the right atrial (ra) and right ventricular (rv) leads resulting in 4 seconds of asystole.The noise was able to be reproduced with isometrics in the bipolar configuration.The physician reprogrammed the device to fixed sensitivity, and the rv sensitivity was adjusted.Additionally, the programming for several of the blanking periods was adjusted.The patient had a remote follow up the following month and there were no new episodes containing noise.The physician plans to continue to monitor.This product remains in service.No adverse patient effects were reported.Additional information received this device was explanted and successfully replaced due to normal battery depletion.The device is returned for analysis.
|