MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS PARIETEX MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS PARIETEX MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number TECR1410DP2
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/28/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, when they were about to place the mesh in the inguinal space on a laparoscopic totally extraperitoneal procedure, they noticed that a 1-2 cm defect was present on the mesh. The surgeon used an alternate mesh to complete the case. There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: a review of the device history has been performed and no failure that may relate to the reported conditions have been noted. The visual examination of the provided picture shows a ¿prosthesis opened and not used/implanted form¿ of the private hospital (b)(6). The mesh was indicated ¿deemed unsuitable when implanted¿. The reported condition could not be confirmed by this examination. No pictures relative to the defect were provided. The reported information is too limited to determine the root cause or most probable cause of the reported incident and/or to determine if the incident is associated with a manufacturing or component discrepancy. Based on our investigation and a complaint history review, the manufacture of the device is not suspected. There is no indication that there is a defective lot or that this event represents an emerging adverse quality trend. If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePARIETEX
Type of DeviceMESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR 01600
Manufacturer (Section G)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR 01600
Manufacturer Contact
lisa hernandez
60 middletown ave.
north haven, CT 06473
2034925563
MDR Report Key8119290
MDR Text Key128872617
Report Number9615742-2018-02630
Device Sequence Number1
Product Code FTL
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K982532
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/04/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received11/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberTECR1410DP2
Device Catalogue NumberTECR1410DP2
Device Lot NumberSSC1149M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-