• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number UNK-P-IPP
Device Problems Inflation Problem (1310); Labelling, Instructions for Use or Training Problem (1318)
Patient Problem Swelling (2091)
Event Date 11/01/2018
Event Type  malfunction  
Event Description
It was reported that the patient experienced auto-inflation with their inflatable penile prosthesis since implant.The device would partially inflate after deflation.The auto-inflation was due to the patient not compressing the pump hard enough to set the valve.The patient's post-operative swelling also added to the appearance of auto-inflation.The issue will resolve over time.Boston scientific has been unable obtain additional information regarding the circumstances surrounding this event to date.Should additional relevant details become available a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer Contact
mike bocchino
10700 bren road w
minnetonka, MN 55343
6515827161
MDR Report Key8122993
MDR Text Key129029859
Report Number2183959-2018-61305
Device Sequence Number1
Product Code FHW
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial
Report Date 12/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUNK-P-IPP
Device Catalogue NumberUNK-P-IPP
Was Device Available for Evaluation? No
Date Manufacturer Received11/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-