Catalog Number H1-M |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/28/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The physician used a hawkone device to treat a severely calcified 60mm lesion with 95% stenosis in the right mid common femoral artery, artery diameter 6mm.
A 6 f non-medtronic sheath and a 0.
014 guidewire were used.
The ifu was followed and vessel was pre-dilated.
Moderate resistance was noted during advancement.
It was reported that when advancing the hawkone for the last pass, the device jumped forward.
The physician then withdrew the device and noticed a failure at catheter and hinge.
A technician advanced cleaning tool down to clean and the tip came completely off.
Flushing was not performed prior to tip separation.
The procedure was completed using a non-medtronic catheter and the vessel was post dilated with a 6/80 in.
Pact admiral.
No patient injury was reported.
|
|
Manufacturer Narrative
|
The device was removed safely from the patient.
Following removal, the proximal part of the hinge looked bent.
No pieces of the cutter were missing.
The patient is doing great and has no issues related to the device failure.
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Device evaluation: visual inspection: the hawkone was removed from the bag and inspected.
It was observed the hawkone was returned in two different segments.
The distal segment was approximately 7cm of the distal assembly.
The housing assembly was fractured off from the unit.
The distal segment showed a fracture at the location where the laser drilled coils start at the proximal end.
The tecothane layer had smooth edges, however the coil within the tecothane was stretched proximally.
A 0.
014" guidewire from the lab was front loaded.
It was verified the guidewire tubing was no torn.
The proximal segment was returned with the distal flush tool over the distal end.
It was observed with the dft, the cutter was visible connected to the driver shaft.
Damage to the cutter window housing segment was noted.
The dft was removed.
It was observed the platinum iridium was crushed and folded in one direction.
A radial fracture of the housing assembly was observed.
The cutter assembly remained intact connected to the drive shaft.
Possible pet from inner housing was noted on the outer rim of the cutter assembly.
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|